Department of Toxicology, University of Würzburg, Würzburg, Germany
(Received May 9, 2005; accepted August 15, 2005)
| Abstract |
|---|
|
|
|---|
2.5 pmol/ml) with or without prior glucuronidase treatment. The results show that care is required for analysis of BPA and its major metabolite BPA-gluc. The LOD obtained and the absence of detectable levels of BPA in samples from individuals suggests that general exposure of humans to BPA is much lower than the worst-case exposure scenario developed.
|
Therefore, sensitive and selective methods to determine BPA in a biological matrix are needed with reduced sample workup to reduce possible contamination of samples with BPA from unknown sources and to quantitate both BPA and BPA-gluc based on stable isotope-labeled standards. The method development also considered the issue of background contamination with BPA present in solvents or released from material used for sample preparation by organic solvents. The combination with a linear ion trap mass spectrometer permits multiple reaction monitoring for quantitation and simultaneous recording of a product ion spectrum for conclusive identification.
| Materials and Methods |
|---|
|
|
|---|
Exposure of Human Subjects to BPA. BPA (25 µg/person) was orally administered in 50 ml of water to three healthy female and three healthy male human subjects (Table 1). All subjects enlisted in the study had to refrain from alcoholic beverages and medicinal drugs 2 days before and throughout the experiment. Subjects did not abuse alcohol and were nonsmokers. Subjects were healthy as judged by medical examination and clinical blood chemistry. The study was carried out according to the Declaration of Helsinki, after approval by the Regional Ethical Committee of the University of Würzburg, Germany, and after written informed consent by the subjects. Urine samples from the subjects were collected in glass vessels at 0, 1, 3, 5, and 7 h after giving BPA. After urine volume was determined, two aliquots (1 ml and 50 ml) were stored at 20°C. Urine and plasma samples from human subjects without known exposure to BPA (age 2650 years) were also collected in glass vessels (urine) or heparinized syringes at random times in the morning or in the afternoon. After urine volume was determined, two aliquots (1 ml and 50 ml) were stored at 20°C. For analysis, part of the plasma samples (100 µl) was diluted with an equal volume of methanol and subjected to centrifugation, and 100 µl of acetonitrile was added, followed by a second centrifugation. From the obtained supernatant, 20 µl were analyzed by LC-MS/MS. Quantitation of creatinine was performed in the 1-ml urine samples separately frozen for creatinine determination as described elsewhere (Amberg et al., 1999
). Glucuronidase treatment of urine and plasma samples was performed as previously described (Völkel et al., 2002
).
|
Identification of BPA-gluc by Information-Dependent Acquisition (IDA). To identify BPA-gluc in urine samples from humans by LC-MS/MS, an IDA method consisting of multiple reaction monitoring (MRM), enhanced resolution (ER), and enhanced product ion (EPI) scan mode was developed. Urine samples (100 µl) were diluted with 100 µl of acetonitrile containing internal standard (d14-BPA-gluc, 42 nM). Samples (10 µl) were separated by a ReproSil-Pur ODS-3 column (5 µm, 150 x 4.6 mm; Maisch, Ammerbuch, Germany) using an Agilent 1100 autosampler and an Agilent 1100 HPLC pump (Agilent Technologies, Waldbronn, Germany). Gradient elution with water (solvent A) and acetonitrile (solvent B) with the following conditions was applied: 0% B for 2 min, followed by a linear gradient linear to 80% B within 20 min. Solvent composition was then kept at 80% B for 2 min. Flow rate was 300 µl/min. IDA experiments were performed on a linear ion trap mass spectrometer (QTrap; Applied Biosystems, Darmstadt, Germany) equipped with a TurboIonSpray source connected to the HPLC system. To record spectral data, a vaporizer temperature of 450°C and a TurboIonSpray voltage of 4.5 kV in the negative ionization mode were applied. Declustering potential (DP) was set to 40 V and N2 was used as collision gas. The MRM mode used transitions from m/z 403 to 227 and from m/z 417 to 241 with a collision energy of 30 V and a collision gas setting of collision-activated dissociation = 4. The ER was performed with a scan rate of 250 amu/s and a fill time of 50 ms. For EPI scans, a scan rate of 4000 amu/s, a collision energy of 30 V, and a fill time of linear ion trap of 400 ms were used. In all experiments DP was set to 40 V. All IDA experiments were performed in the range of m/z 50 to m/z 420 with an entrance potential of 8 V and enabled Q0-trapping.
Quantitation of BPA-gluc. All quantitations were performed using a triple quadrupole instrument (API 3000; Applied Biosystems) equipped with a TurboIonSpray source. The HPLC parameters used were identical to those described above for identification. The HPLC system was directly coupled to the mass spectrometer. Analytes were detected in the negative-ion mode at a vaporizer temperature of 400°C and a TurboIonSpray voltage of 4.0 kV. Spectral data were recorded with N2 as collision gas (collision-activated dissociation = 4) in the MRM mode with a dwell time of 500 ms for each transition monitoring the MS/MS ion transitions shown in Table 1. Quantitation of BPA-gluc was based on calibration curves obtained after addition of known amounts of BPA-gluc (0 nM, 20 nM, 39 nM, 78 nM, 157 nM, and 314 nM) to urine samples from the subjects collected before the exposure. Calibration curves were constructed from six data points using Analyst 1.3.1 (Applied Biosystems) with R2 > 0.99. Intra-assay variability was below 15%. Acquisition conditions for mass spectra are detailed in Table 2. Under these conditions, calibration curves for BPA-gluc were linear up to concentrations of 1900 nM with R2 = 0.998 and an accuracy between 88% and 115% (Table 3).
|
|
Quantitation of BPA. Samples both before and after glucuronidase treatment were analyzed by LC-MS/MS using the API 3000. Separations were performed with water (A) and acetonitrile (B) as solvents: 60% B for 2 min, followed by a linear gradient to 80% B within 20 min. The solvent composition was then kept at 80% B for 2 min; flow rate was 300 µl/min. All other parameters were identical to those described for the analysis of BPA-gluc. Quantitation of BPA was also performed relative to the stable isotope-labeled internal standard d16-BPA (Table 2). With this method, calibration curves for BPA were linear up to concentrations of 772 nM with R2 = 0.998 and an accuracy between 95% and 108% (Table 3).
| Results |
|---|
|
|
|---|
The developed methods apply stable isotope-labeled standards as a basis for quantitation of both BPA and BPA-gluc and LC-MS/MS with negative ionization, and were optimized for both analytes for optimal sensitivity (Tables 2 and 3). Under these conditions, both BPA and BPA-gluc are deprotonated to give [M H] (227.0 amu) for BPA, respectively [M H] (403.2 amu) for BPA-gluc. Addition of buffer salts or acid to improve separation was not possible since the addition of buffer salts or acids to the HPLC solvents dramatically reduced sensitivity. The use of stable isotope-labeled internal standards permitted a clear identification of the signals for both BPA-gluc and BPA despite the small shifts in retention time observed in the absence of buffers (Table 3). Moreover, collision-induced fragmentation results in characteristic product ions from the precursor ion at 403.2 amu of BPA-gluc (Table 2). These fragmentation patterns were included in the acquisition method, and quantitations of BPA and BPA-gluc are reported relative to the internal standards.
Two different HPLC separation methods were developed. The first method provided a separation of BPA and BPA-gluc and a separation of BPA-gluc from the solvent front. This method was initially intended for the simultaneous quantitation of both BPA and BPA-gluc in one separation. A clear separation of BPA-gluc from the salts and other components of urine or plasma, which elute close to the solvent front, is necessary since these components may suppress ionization of BPA-gluc and thus reduce sensitivity. However, using this system, peaks indicative of the presence of BPA at the retention of authentic BPA (Kuch and Ballschmiter, 2001
; Völkel et al., 2002
) were observed when injecting only solvents (water/acetonitrile, 1:1 v/v), or even when the gradient was started without sample injection (Fig. 2, A and B). Integration of peak areas observed suggested the presence of concentrations up to 25 nmol of BPA/ml of urine or plasma. The peaks representing BPA increased over time after repeated injection of purified solvent. These observations suggest a contamination of the HPLC solvents with BPA, a leaching of BPA from plastic equipment used, and/or adsorption of BPA to the column material. Change or cleanup of solvents by distillation, use of different HPLC columns, or minor modifications of the gradient did not completely solve the problem of BPA background. Sample workup exclusively in purified glass vessels without contact to plastics somewhat reduced the suggested concentrations of BPA but was also not able to reduce the BPA contents in blanks to below the LOD.
|
Therefore, separate methods were developed to quantify BPA-gluc and BPA. The method for BPA-gluc used the gradient starting with a high water content of the mobile phase and permitted the quantitation of BPA-gluc with a limit of detection of 0.9 pmol/ml when the standard was analyzed after dilution in purified water. However, despite the use of a specific analytical method including HPLC separation, stable isotope internal standard, and presumably selective MS/MS transition, several peaks were observed in the region where BPA-gluc elutes from the column when analyzing urine samples from humans without intentional exposure to BPA (Fig. 3, A and B). Although different HPLC columns (Synergi Hydro 150 x 2 mm; 80 A, 4 µm; Nucleosil C18 150 x 2 mm, 100 A, 5 µm; ReproSil-Pur C18-AQ 150 x 4.6 mm, 120 A, 3 µm) were used, HPLC solvents were acidified, and gradient parameters were changed, the best resolution of these components and spiked BPA-gluc obtained is shown in Fig. 3B. The presence of these interfering compounds reduced the LOQ for BPA-gluc to 65 pmol/ml [signal to noise ratio (S/N) 8] in urine, due to a 50-fold reduced S/N. Addition of BPA-gluc (final concentration 500 pmol BPA-gluc/ml) to a urine sample resulted in a defined peak with an S/N of 31 presented in Fig. 3C. The peaks at retention times 8.85, 9.09, and 9.18 min presumably represent isobaric glucuronides that show identical fragmentation.
|
In urine samples from humans administered BPA, a peak with the mass spectral information supposed to be characteristic for BPA-gluc eluted at the expected retention time (Fig. 3D). However, due to the closely eluting peaks of other compounds, confirmation of the identity of BPA-gluc was required, and the presence of presumably present isobaric glucuronides indicates a need for a more conclusive identification of BPA-gluc to avoid false-positive results. Therefore, urine samples from individuals given BPA were also analyzed by a LC-MS/MS QTrap instrument using the IDA scan method to confirm the presence of BPA-gluc. The linear ion trap mass spectrometer combined with the software tool IDA allows mass scans including a MRM for both BPA-gluc and the internal standard d14-BPA-gluc (no clear separation of BPA-gluc and d14-BPA-gluc was achieved) to select the analytes, an enhanced resolution to record higher resolved [M H] spectra, and an enhanced product ion scan to record a mass spectrum of each analyte recognized by the MRM mode. These mass spectral data are obtained in a single separation. In Fig. 4A, the characteristic EPI mass spectrum of BPA-gluc is presented, and Fig. 4B shows the spectrum of BPA-gluc obtained by IDA analysis in a urine sample collected 3 h after application of 25 µg of BPA to a human subject. If all EPI spectra of m/z 403 (BPA-gluc) and m/z 417 (d14-BPA-gluc) recorded within the peak width of BPA-gluc are summed up, the resulting EPI mass spectrum gives a typical fragment (m/z 241) of d14-BPA-gluc due to coelution and identical collision energy for both glucuronides (Fig. 4B). In addition, the signal m/z 113 formed from both precursors (m/z 403 and m/z 417) represents the sum of both precursors and is therefore higher compared with the spectrum of the precursor m/z 403 alone (Fig. 4A). In urine and calibration samples in water and acetonitrile (50:50 v/v) without the addition of the internal standard d14-BPA-gluc, the fragment of m/z 241 was not observed and the presumed isobaric glucuronides showed more or less changed mass spectra (data not shown). The combination of both LC-MS/MS methods permitted a conclusive identification of BPA-gluc even in low concentrations. These results are more reliable as compared with methods that only use single chromatographic traces and retention time.
|
Analysis of Urine Samples for BPA-gluc and BPA after Oral Administration of BPA to Human Subjects. The two procedures were applied to analyze samples from an application of BPA to human subjects in a single oral dose in the range of the worst-case assumption for daily exposure. All samples were analyzed independently with both HPLC systems, and with and without glucuronidase treatment. This combination of analytical methods gives information on the presence of both BPA and BPA-gluc and on the total content of BPA in a sample. The excretion of BPA and BPA-gluc was quantitated over time in urine samples, but BPA-gluc was only quantified in samples in which its presence could be confirmed by EPI-mass spectrometry (time points 1, 3, and 5 h). In male subjects, 93 ± 19 nmol (85% of the applied BPA dose) and in female subjects, 83 ± 16 nmol (75% of the applied dose) were recovered as BPA-gluc (Fig. 5) within 5 h after application. An elimination half-life of approximately 4 h was calculated from the excretion rates. The half-life and the recovery is similar to that reported in a previous study (Völkel et al., 2002
), which used higher doses of d16-BPA applied to human subjects. The highest concentrations of BPA-gluc from 221 to 611 pmol/mg creatinine and 117 to 345 pmol/mg creatinine were observed in urine samples collected 1 h and 3 h, respectively, after administration of 25 µg of BPA/subject (Fig. 5). In plasma samples from this exposure, BPA-gluc could also be detected due to the absence of interfering peaks (data not shown).
|
|
BPA and BPA-gluc in Urine and Plasma Samples from Human Subjects without Intentional Exposure to BPA. Urine and plasma samples from human subjects (7 males and 12 females) were randomly collected between 10 AM and 7 PM on different days and analyzed for BPA-gluc and BPA. In most of the urine samples, small peaks with the characteristic MS/MS transitions for BPA-gluc were detected at retention times very close to those of authentic BPA-gluc, but the concentrations were lower than 65 pmol/ml (LOQ), corresponding to <65 nmol of BPA-gluc in an average urine volume of 1 liter. Due to the low concentrations and the presence of other analytes with identical mass spectral characteristics, an unequivocal identification of BPA-gluc with IDA-MRM-ER-EPI was not possible. After glucuronidase cleavage, the BPA concentrations determined in urine were always below the LOD of 1.14 µg/l (5 pmol/ml), suggesting the presence of very low concentrations of BPA, at best. However, addition of even small concentrations of BPA or BPA-gluc to the samples gave definite responses (Figs. 6C and 7). None of the plasma samples analyzed (both with and without glucuronidase treatment) contained detectable concentrations of BPA or BPA-gluc, also indicating plasma concentrations of BPA below 0.5 µg/l.
|
| Discussion |
|---|
|
|
|---|
BPA-gluc was analyzed after fragmentation of the [M H] anion formed in the negative ionization mode of the electrospray source (Inoue et al., 2002
; Völkel et al., 2002
). With this technique, BPAgluc may be identified by monitoring specific mass transitions. However, since the method included a loss specific for glucuronides, presumed isobaric glucuronides of unknown structures eluted closely to BPA-gluc. The fragmentation of these compounds gave identical spectral information (fragment with m/z 113 and [M H] of the glucuronide moiety).
A linear ion trap mass spectrometer was, therefore, additionally used to confirm the presence of BPA-gluc in low concentrations. With this instrument, an EPI spectrum can be recorded simultaneously to the MS/MS transitions specific for BPA-gluc (Fig. 4). The detection of the M-glucuronide fragment of d14-BPA-gluc in these spectra, not seen in samples without d14-BPA-gluc, serves as additional evidence for the presence of BPA-gluc. Application of this combination of mass spectrometry procedures clearly confirmed that, after controlled exposure to low doses of BPA, BPA-gluc is excreted as a metabolite of BPA.
However, the method using the triple quadrupole was not sufficiently selective to be applied to samples from human subjects without intentional exposure to BPA. Therefore, the strategy was changed to determine free BPA in urine or blood samples both with and without glucuronidase cleavage. BPA may be quantified with high sensitivity by LC-MS/MS; however, issues with background contamination needed to be solved before the method could be applied. Likely, water and/or acetonitrile used as HPLC solvents contain traces of BPA (Kuch and Ballschmiter, 2001
; Völkel et al., 2002
). Another source for contamination may be the release of BPA from plastics by organic solvents (Krishnan et al., 1993
; Hutson, 1998
; Fung et al., 2000
). Indeed, sample workup with glassware gave a trend to lower BPA background levels. However, only a major change in HPLC separation conditions gave BPA concentrations below the limit of detection when solvent samples were injected. This could be explained by a trapping effect of the HPLC column for BPA at lower concentrations of organic solvents. Small amounts of BPA presumably contained in the HPLC-grade water may be trapped by the column; after reaching higher concentrations of acetonitrile in the eluent, the trapped BPA elutes from the column (Meyer, 1999
).
Based on these observations, the optimized analytical procedure for exposure monitoring of BPA was considered to be the analysis of free BPA, both with and without prior hydrolysis of BPA-gluc, to determine free BPA and total BPA. The content of free BPA is relevant in the context of possible effect concentrations of BPA, since some effects of low-dose BPA administration (>20 µg/kg b.wt. per day) have been described (vom Saal et al., 1997
; Takahashi and Oishi, 2000
), and the blood levels of free BPA are relevant for assessment of endocrine activity since BPA-gluc is devoid of estrogenicity (Matthews et al., 2001
).
When applying this procedure to a number of randomly collected blood and urine samples from human subjects without intentional exposure to BPA, the content of BPA in all samples was below the LOD of 5 pmol/ml (
1.14 µg/l), suggesting that the human exposure to BPA from diffuse sources in the environment is low. The levels observed in European subjects are well below those reported previously for populations in Korea and Japan, where mean values of 2.8 µg/l and 9.5 µg/l, respectively, were found in urine (Ouchi and Watanabe, 2002
; Kim et al., 2003
; Yang et al., 2003
). In recent studies in the United States, mean urinary concentrations of BPA were 1.33 µg/l determined by GC-MS after glucuronidase treatment (Calafat et al., 2005
). In Japan, daily excretion of BPA with urine was reported as 1.2 µg/day (mean) using GC-MS/MS (Arakawa et al., 2004
). The data presented here suggest that BPA concentrations in human urine in Europe are, at best, in a similar range.
Due to the rapid and complete elimination of BPA in humans as BPA-gluc in urine (Völkel et al., 2002
), the determined concentration of BPA in urine suggests that the human exposure to BPA is less than 2.3 µg/person/day (based on a volume of 2 liters of urine excreted). This value corresponds to a BPA dose of 38 ng/kg b.wt. per day for a 60-kg adult and, thus, is well below the daily doses of BPA that have sometimes been reported to cause responses of unknown toxicological relevance in highly sensitive animal systems (20 µg/kg/day) (Timms et al., 2005
).
The lack of detection of BPA in plasma with a LOD of 0.5 ng/ml also suggests that the plasma levels of BPA and BPA-gluc, respectively, in Europeans are below the BPA plasma concentrations reported in nonobese Japanese women (0.71 ± 0.09 ng/ml as analyzed by enzyme-linked immunosorbent assay) (Takeuchi et al., 2004
) and lower than the previously reported median of 3 ng of BPA/ml of plasma in pregnant females in Europe (Schönfelder et al., 2002
).
The capacity of the procedure to detect low levels of BPA and BPA-gluc in samples from human subjects administered a low dose of BPA, which represents the estimated maximum human exposure (EU Report, 2003
), showed the capability of the procedure for exposure monitoring and also confirms previous observations on the rapid and almost complete excretion of BPA as BPA-gluc within a few hours after oral exposure. The rapid elimination of BPA as seen in this and a previous study of human excretion kinetics of BPA also indicated a need for a very frequent monitoring of BPA in blood and urine, since BPA taken up from food will be rapidly eliminated and will only be present in detectable concentrations for a relatively short period.
In summary, the results show that exposure monitoring to potentially ubiquitous chemicals requires a very detailed approach to account for possible interference by other chemicals present in the media and also requires specific analytical methods such as MS/MS. However, even when using these specific methods, structure confirmation is required. Therefore, methods to quantify BPA or BPA-gluc by HPLC-UV, HPLC-FLD, or HPLC with electrochemical detection or after fluorophore derivatization in urine of nonintentionally exposed humans may have only limited applicability (Ouchi and Watanabe, 2002
; Takeuchi and Tsutsumi, 2002
; Matsumoto et al., 2003
; Mao et al., 2004
; Kawaguchi et al., 2005
), since the use of internal standards with these methods is difficult and many possibly interfering compounds with fluorophores or electrochemical potential are present in biological samples. Quantitation of parent BPA with GC-MS has a sufficient sensitivity and a higher power of separation as compared with HPLC (Schönfelder et al., 2002
), but the procedure requires solvent extraction and/or further sample preparation as a possible source of contamination with BPA and needs to be well evaluated to avoid production of false-positive results. GC-MS analysis for BPA in human blood or urine samples also needs to account for the fact that, due to the rapid conjugation of BPA to BPA-gluc, most of the BPA is present as BPA-gluc. Therefore, a separation of BPA-gluc and free BPA must be achieved before extraction and derivatization, since the derivatization at higher temperature may hydrolyze BPA-gluc and overestimate the level of free BPA. Therefore, a LC-MS/MS method is more reliable and permits the quantitation of both free and conjugated BPA, which is relevant for assessment of postulated human health risks due to low-dose BPA exposures.
| Footnotes |
|---|
Article, publication date, and citation information can be found at http://dmd.aspetjournals.org.
ABBREVIATIONS: BPA, bisphenol A; BPA-gluc, BPA-monoglucuronide; DP, declustering potential; EPI, enhanced product ion; ER, enhanced resolution; IDA, information-dependent acquisition; LOD, limit of detection; LOQ, limit of quantitation; MRM, multiple reaction monitoring; S/N, signal to noise ratio; GC-MS, gas chromatography-mass spectrometry; LC-MS/MS, liquid chromatography-tandem mass spectrometry; HPLC, high-performance liquid chromatography; amu, atomic mass unit(s).
Address correspondence to: Dr. W. Dekant, Department of Toxicology, University of Würzburg, Versbacherstrasse 9, 97078 Würzburg, Germany. E-mail: dekant{at}toxi.uni-wuerzburg.de
| References |
|---|
|
|
|---|
and ß. Chem Res Toxicol 14: 149157.[CrossRef][Medline]
This article has been cited by other articles:
![]() |
U. M. D. Schauer, W. Volkel, and W. Dekant Toxicokinetics of Tetrabromobisphenol A in Humans and Rats after Oral Administration Toxicol. Sci., May 1, 2006; 91(1): 49 - 58. [Abstract] [Full Text] [PDF] |
||||
| |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||