TABLE 1

Accurate mass measurement of the fragments after CID of a flutamide standard solution (18 μM) For LC and MS settings, see Materials and Methods.


Fragment

Experimental [M – H]

Theoretical [M – H]

Difference

Fragment Loss
m/z mTh
A 275.0650a 275.0643 0.7
B 229.0598b 229.0714 –11.6 A – NO2
C 205.0223b 205.0225 –0.2 A – C4H6O
D 202.0090b 202.0116 –2.6 A – C3H7, NO
E 186.0145c 186.0167 –2.2 A – C3H7, NO2
F 182.0041c 182.0054 –1.3 D – HF
G
154.0099c
154.0183
–8.4
F – C=O
  • a Lock mass 276.7987 used

  • b Lock mass 203.0821 used

  • c Lock mass 179.0345 used