TABLE 2

Intraday and interday validation of P-gp synthetic proteotypic peptide quantification

The synthetic proteotypic peptide was spiked to the membrane fraction of a membrane vesicle preparation originating from untransfected HEK293 cells (matrix) at two concentrations (100 and 1000 pmol/l). Preparations were performed at three independent occasions.

ExperimentMatrix100 pM1000 pM
Day 1
    n33
    Mean50.891.21017.6
    Accuracy (% RE)−8.81.8
    Precision (% CV)9.610.9
Day 2
    n33
    Mean41.6113.1935.7
    Accuracy (% RE)13.1−6.4
    Precision (% CV)12.18.5
Day 3
    n33
    Mean49.187.4918.5
    Accuracy (% RE)−12.6−8.1
    Precision (% CV)11.310.2
Interday
    n99
    Mean97.2957.3
    Accuracy (% RE)−2.8−4.2
    Precision (% CV)12.010.7
  • %RE, percentage relative error; %CV, coefficient of variation.