TABLE 1

Comparison of features of mass cytometry and flow cytometry

FeatureFlow CytometryMass Cytometry
No. of biomeasures that can be multiplexed3–18 ParametersUp to 40 parameters (the instrument is capable of multiplexing 100 biomeasures)
CompensationFluorescence compensation absolutely requiredNo fluorescence-style compensation required. However, correction for isotopic impurities in metal tags and oxides is needed
Status of cellsBoth live and fixed cells can be analyzedCells are required to be fixed prior to analysis
Cell sortingYes (in sorters)No
Efficiency of analysis of injected samples>95%30–40% of injected cells are analyzed
Speed of acquisitionUp to 60,000 cells/sUp to 1000 cells/s
High-throughput compatibleYes (with a 96-well format)Yes (with a 96-well format)