Summary

There are some basic requirements for air conditioning systems in operating theatres.

  1. 1.

    Obviously the regulation of temperature and humidity is necessary, but cleaning of fresh air and circulating air is much more important.

  2. 2.

    Concerning cleaning, a high standard of filtration of air is demanded to avoid contamination of wounds.

  3. 3.

    So far only HEPA-filters, a technical description, give security and safety. They filter particles down to under 1 μm and have a filtration rate of more than 99.9%.

  4. 4.

    The ventilation rate is directly related to the germ concentration in the air. The higher the ventilation rate the cleaner the air during working.

  5. 5.

    To avoid turbulences unidirectional or uniform flow is necessary. This technique has become popular as the laminar flow technique. To get such a flow an air flow velocity of 0 45±0.1 m/s is necessary.

  6. 6.

    The arrangement of the different working places in an operating theatre is very important. Instrument tables and the operation field itself have to be in the zone of first air.

  7. 7.

    One of the commonest germ dispersion sources is the human body. Proper dressing and behaviour have to be observed. The airflow behind a person (in laminar cross flow) or under a person (in laminar down flow for instance when the neurosurgeon bends his head over the operating field) has to be regarded as contaminated.

  8. 8.

    Hygienic controls must be made in the operation theatre when unused and also under working conditions, in a sophisticated manner.

If all these points are respected a real clean room technique is possible with considerable lowering of the risk of infection for the patient. Obviously all the other rules of hygiene,i.e., asepsis and antisepsis, elaborated since the days of Lister, have to be strictly respected.

This is a preview of subscription content, log in via an institution to check access.

Access this article

Price excludes VAT (USA)
Tax calculation will be finalised during checkout.

Instant access to the full article PDF.

Institutional subscriptions

Similar content being viewed by others

References

  • Allander, C., Abel, E., Untersuchung eines neuen Einblasesystems für Operationsräume. Dechema-Monographie Bd.69 (1968), 297–306.

    Google Scholar 

  • Austin, P. R., Timmerman, S. W., Design and operation of clean rooms. Detroit: Business News Publ. 1965.

    Google Scholar 

  • Botzenhardt, K., Rüden, H., Zur Beurteilung von Klimaanlagen im Krankenhaus. Klima-Kälte-Technik7 (1973), 139–143.

    Google Scholar 

  • —, Wanner, H. U., Technik des Keimnachweises in Laminar-flow Operationsräumen. In: Sattel, W., Peiper, H. J., Reinraumtechnik, Anwendung in der Medizin, 81–88. Berlin-Heidelberg-New York: Springer. 1977.

    Google Scholar 

  • Charnley, J., A clean-air operating enclose. Brit. J. Surg.51 (1964), 202.

    PubMed  Google Scholar 

  • Nouri, Z., Reinfeldverfahren zur Erzeugung lokaler luftkeimarmer Bereiche. Bericht anläßlich der 5. Fachtagung für Krankenhaustechnik der Medizinischen Hochschule Hannover, April 1977 (gedruckte Fassung 1978, 41–49).

  • Peiper, H. J., Sattel, W., Schubert, E, Verminderung der Infektionsgefahr durch Anwendung einer turbulenzarmen Verdrängungsströmung im Operationssaal. Der Chirurg45 (1974), 545–549.

    Google Scholar 

  • Regenscheit, B., Die Archimedes-Zahl-Nennzahl zur Beurteilung von Luftströmungen. Gesundh.-Ing.91 (1970), 172.

    Google Scholar 

  • Russenberger, H. J., Wanner, H. U., Keimgehalt der Raumluft in Abhängigkeit des Luftwechsels. Zbl. Bakt. Hyg. I. Abt. Orig. A.227 (1974), 564–567.

    Google Scholar 

  • Sattel, W., Peiper, H. J., Reinraumtechnik. Anwendung in der Medizin. Berlin-Heidelberg-New York: Springer. 1977.

    Google Scholar 

  • — —, Berendt, R., Auswirkung auf die Organisation in LF-Räumen und LF-Kabinen. In: Sattel, W., Peiper, H. J., Reinraumtechnik, Anwendung in der Medizin, 71–80. Berlin-Heidelberg-New York: Springer. 1977.

    Google Scholar 

  • Schepelmann, F., Greiner, L., Pia, H. W., Complications following of herniated lumbar discs. In: Advances in neurosurgery 4 (Wüllenweber, R.,et al., ed.), pp. 52–54. Berlin-Heidelberg-New York: Springer. 1977.

    Google Scholar 

  • Schicht, H. H., Grundlagen und Anwendungen der Reinraumtechnik. Der Gesundheitsingenieur93 (1972), 232–241.

    Google Scholar 

  • —, Steiner, W., Klimatechnische Beiträge zur Asepsis im Operationssaal. Technische Rundschau4 (1972), 2–14.

    Google Scholar 

  • Schütz, H., Probleme der Luftreinhaltung und -hygiene im Krankenhaus. Klima-Technik4 (1970), 12–27.

    Google Scholar 

  • Skiba, H., Klimatisierung von Op-Räumen mit Hilfe der Reinraumtechnik. Sanitär- und Heizungstechnik1 (1975), 19–24;2 (1975), 66–72.

    Google Scholar 

  • Stokes, E. A., Mulcaster, K. D., Air filtration. Steam and Heating Engineer10 (1969), 3.

    Google Scholar 

  • Strauss, H. J., Strömungstechnische und luftfiltertechnische Grundlagen der LF-Reinraumtechnik. In: Sattel, W., Peiper, H. J., Reinraumtechnik, Anwendung in der Medizin, 1–33. Berlin-Heidelberg-New York: Springer. 1977.

    Google Scholar 

  • Thomsen, K., Krebs, D., Fehlerhafte Klimaanlage infizierte Operationstrakt. Deutsches Ärzteblatt Heft10 (1972), 544.

    Google Scholar 

  • Wanner, H. U., Bakteriologische Kontamination in konventionell belüfteten Operationssälen. Berichte Unfallmedizinische Arbeitstagung, Göttingen: 1972.

  • - Möglichkeiten und Grenzen der Reinraumtechnik in der Medizin. Chemische Rundschau 53–56 (1974).

  • Weber, B. G., Stühmer, G., Meierhans, L., Sterile Operationsboxen. Z. Orthop.109 (1971), 803.

    PubMed  Google Scholar 

  • — — —, Die Bekleidung des Operationsteams in einer sterilen Operationskabine. In: Internationales Symposium für Reinraumtechnik. Reinraumtechnik I, 132, Solothurn: Vogt-Schild. 1973.

    Google Scholar 

  • Whitcomb, C., Technical lecture series on industry and microbial contamination control. Palo Alto/Calif.: Stanford University. 1969.

    Google Scholar 

  • Whitfield, W. J., A new approach to clean-room design. Reprint No. SC-4673 (RR). Albuquerque/New Mexico: Sandia Corp. 1962.

    Google Scholar 

  • Winkler, S., Roemer, G. B., Matz, K., Konstruktion und Wertung der Klimaanlagen verbessern. Deutsches Ärzteblatt Heft10 (1972), 548.

    Google Scholar 

Download references

Author information

Authors and Affiliations

Authors

Rights and permissions

Reprints and permissions

About this article

Cite this article

Lorenz, R. Air conditioning systems. Acta neurochir 55, 49–61 (1980). https://doi.org/10.1007/BF01808920

Download citation

  • Issue Date:

  • DOI: https://doi.org/10.1007/BF01808920

Keywords

Navigation