Communication
General implementation of the ERETIC™ method for pulsed field gradient probe heads

https://doi.org/10.1016/j.jmr.2008.07.013Get rights and content

Abstract

A capacitive coupling between a secondary radiofrequency (rf) channel and the gradient coil of a standard commercially available high resolution NMR spectrometer and probe head is described and used to introduce a low level exponentially damped rf signal near the frequency of the primary rf channel to serve as an external concentration standard, in analogy to the so-called ERETIC™ method. The stability of this inexpensive and simple to implement method, here referred to as the Pulse Into the Gradient (PIG) approach, is superb over a 14-h period and both gradient tailored water suppression and one-dimensional imaging applications are provided. Since the low level signal is introduced via the pulsed field gradient coil, the coupling is identical to that for a free induction signal and thus the method proves to be immune (within 5%) to sample ionic strength effects up to the 2 M NaCl solutions explored here.

Introduction

It is well known that nuclear magnetic resonance (NMR) spectroscopy is quantitative, i.e. measured peak intensities directly correspond to nuclear spin concentrations [1]. This spectral intensity dependence combined with chemical shift and scalar J coupling resolution make NMR spectroscopy a powerful tool since the relative number of the different chemical environments and their geometric relationship in a molecule can be determined from the NMR spectrum [2]. The absolute nuclear spin concentration within a given molecular environment or at a given chemical shift cannot be determined directly from the spectrum without exhaustive spectrometer calibration or the inclusion of an internal standard with known concentration. Provided that a “friendly” internal standard can be found, i.e. a compound that does not react with, exists in the same phase as, and has a chemical shift different than the sample being studied or the analyte, the problem of absolute quantitation is solved. Here the ratio of the integrated peak intensity of the analyte to that for the standard multiplied by the standard concentration gives the analyte concentration. However, there are many situations where an internal standard cannot be used [3], [4], [5] and external referencing is required. One alternative approach that uses an electronic reference (ERETIC™) has become an increasingly popular option [6]. The ERETIC™ method uses a low voltage, exponentially damped, synthetic radiofrequency (rf) signal near the Larmor frequency of the analyte to effectively provide an external concentration standard with the main advantage being that the corresponding signal intensity is a faithful representation of electrical variations in the NMR detection circuitry. While all ERETIC™ implementations require either an additional spectrometer channel or a high frequency waveform generator clocked with the NMR instrument, two distinct modes of coupling the ERETIC™ signal into the NMR detection electronics have been presented in liquids and solids NMR as well as magnetic resonance imaging [6], [7], [8], [9], [10], [11], [12], [13]. However, both approaches have their limitations. Historically, the first ERETIC™ applications coupled the low voltage calibration signal into the NMR detection electronics with a broadband untuned coil, an approach most efficiently and reproducibly accomplished by mounting the ERETIC™ antenna inside of the NMR probe head close to the rf pulsing/receiving coil [6], [7], [10]. Drawbacks to this approach are obvious. The NMR probe head must be customized to accommodate the broadcast antenna, an option that is not viable for most commercial NMR instrumentation, and the antenna also behaves as a receiver picking up high power rf pulses, an effect that at best decreases probe head efficiency and at worst damages the ERETIC™ signal generator. Both of these problems have been circumvented in more recent ERETIC™ applications, which use a directional coupler on the second channel of a double resonance NMR spectrometer and probe head to present the low voltage ERETIC™ signal to the sample and primary detection channel [8], [9], [11]. In this way the tuned LC circuit for one frequency is used to broadcast the ERETIC™ signal at another frequency. Although no NMR probe modifications are required, provided that a double resonance probe is used, in practice the method is extremely sensitive to the mutual coupling between rf channels, an effect that is controlled by the respective tuning and matching capacitors on both channels. In addition, as the rf isolation between channels improves, the performance of ERETIC™ drops. In spite of the concentration quantitation offered by the ERETIC™ method the implementation awkwardness described above limits its broad application, and thus alternative external referencing methods have been proposed [14].

This work describes a straightforward implementation of the ERETIC™ method that can be quickly integrated into standard spectrometers with the minimal addition of extra hardware. This scheme combines the positive aspects of both approaches described above to present the ERETIC™ signal to the detection electronics with only the addition of a simple capacitor. The method is completely general and can be implemented on any probe head equipped with a pulsed field gradient (PFG) coil, either in high-resolution or imaging setups, allowing the untuned PFG coil available in virtually all modern liquids or solids NMR probe heads to serve as an ERETIC™ antenna. An rf directional coupler is not needed as the ERETIC™ signal is capacitively coupled into the cable connecting the gradient amplifier to the gradient channel in the NMR probe head as shown in Fig. 1. The capacitive coupling of this method, hereafter referred to as the Pulse Into the Gradient (PIG) approach, has the benefit of electrically isolating the gradient channel from the low voltage ERETIC™ signal source, i.e. the long time scale μs–ms high current DC field gradient pulses do not feed back into the ERETIC™ channel, but the higher frequency ns–μs ERETIC™ signal couples into the gradient cable. The use of the gradient circuit included in the probe head offers several advantages in comparison to existing ERETIC™ signal coupling strategies. Since the PFG coil is not tuned, a flat frequency response is observed and the extremely small mismatch between rf and gradient coil positions along with their close proximity is enough to inductively couple the ERETIC™ signal into the detection electronics. The use of the untuned PFG coil as an ERETIC™ antenna guarantees the possibility of extending the method to very high field and thus Larmor frequencies of up to a few GHz. An additional attractive feature of the PIG approach is the concomitant increase in ERETIC™ signal stability offered by the properly shielded PFG coil and cables.

Section snippets

Experimental

All deuterated compounds were obtained from Cambridge Isotope Laboratories, while all other chemicals were obtained from Sigma–Aldrich and used without further purification. NMR spectra were collected using a 400 MHz Bruker Avance spectrometer equipped with two distinct commercial 5 mm z gradient Bruker probe heads. Specifically, the stability of the ERETIC™ signal was investigated using both a direct triple-resonance X/1H/31P TBO and an inverse 1H/X BBI probe heads. All other experiments

Results and discussion

As can be gathered from Fig. 1, the critical elements in this external signal referencing approach are the PIG box capacitors which are used to pass the high frequency low voltage ERETIC™ signal but block the high power DC field gradient pulses from reaching and damaging the ERETIC™ signal source, here the F3 spectrometer channel. The 18 mA current break through the capacitor when a 14.8 A 10 ms square DC gradient pulse is applied provides ca. 135 dB isolation between the gradient and F3 channel.

Conclusion

The primary goal of this work was to introduce a simple, practical, and inexpensive modification to a modern high resolution NMR spectrometer that permits external sample concentration referencing. The approach couples an ERETIC™ reference signal through a capacitor into the cable connecting the DC PFG amplifier to the PFG coil contained in most commercially available high resolution liquid and solid state NMR probe heads. The stability of the Pulse Into the Gradient (PIG) approach is superb

Acknowledgments

M.P.A. is a David and Lucile Packard and Alfred P. Sloan foundation fellow. Support of M.P.A. as a visiting scholar by ISM2 from the Aix-Marseille Université and of D.N.S. by the UC Davis International Programs Office is gratefully acknowledged. Thanks are also due to Spectropole for providing access to the NMR facilities of the Fédération des Sciences Chimiques de Marseille.

References (15)

  • A. Abragam

    The Principles of Nuclear Magnetism

    (1961)
  • R.R. Ernst et al.

    Principles of Nuclear Magnetic Resonance in One and Two Dimensions

    (1987)
  • P.T. Callaghan

    Principles of Nuclear Magnetic Resonance Microscopy

    (1991)
  • L. Shintu et al.

    Is high-resolution magic angle spinning NMR a practical speciation tool for cheese samples? Parmigiano Reggiano as a case study

    Magn. Reson. Chem.

    (2004)
  • D.N. Sobieski et al.

    Towards rapid throughput NMR studies of full wine bottles

    Solid State Nucl. Magn. Reson.

    (2006)
  • L. Barantin et al.

    A new method for absolute quantitation MRS metabolites

    Magn. Reson. Med.

    (1997)
  • S. Akoka et al.

    Improvement of the ERETIC method by digital synthesis of the signal and addition of a broadband antenna inside the NMR probe

    Instrum. Sci. Technol.

    (2002)
There are more references available in the full text version of this article.

Cited by (0)

View full text