Determination of clozapine and desmethylclozapine in human plasma by high-performance liquid chromatography with ultraviolet detection

J Chromatogr. 1993 Mar 5;613(1):168-73. doi: 10.1016/0378-4347(93)80212-m.

Abstract

A method using reversed-phase high-performance liquid chromatography for the simultaneous determination of clozapine and its desmethyl metabolite in human plasma has been established. Clozapine and N-desmethylclozapine were extracted with n-hexane-isoamyl alcohol (98.5:1.5, v/v). Protriptyline served as the internal standard. The limits of detection for clozapine and desmethylclozapine are 2 and 1 ng/ml, respectively. The sensitivity and precision of this method can be utilized for pharmacokinetic studies and therapeutic drug monitoring regimens.